Lovibond® MD 100 Benchtop COD Colorimeter Set-up

COD Colorimeter Set-up, Benchtop, COD HR TT, COD LR TT, COD MR TT Test Name, 200 to 15000 mg/L COD, 3 to 150 mg/L COD, 20 to 1500 mg/L COD Measuring Range, Dichromate/H2SO4 Chemical Method, For Raw Water Treatment, Waste Water Treatment

  • Manufacturer Part No: 276130
  • SECO Part No: 17-2480-31
  • Sold As:
  • Availability: loading

The Lovibond® COD setups allow highly sensitive and precise water testing with minimum effort. They measure the ST-COD concentration by photometric detection employing a linear relationship between absorbance and concentration. After adding the sample to a Lovibond® COD tube test (LR, MR according to ISO 15705:2002), it is heated in the reactor for 2 hr at 150°C and then analyzed in the photometer. The COD setups comprise the photometer, 25 tube tests for each of the two lower measuring ranges, a reactor for sample digestion and a vial stand.

  • Determination of the chemical oxygen demand index (ST-COD)
  • ST-COD sealed tubes ready-for-use
  • Suppression of chloride interference upto 1000 mg/L (LR and MR) upto 10000 mg/L (HR)
  • Mercury-free tube tests, in absence of chloride interference
  • Four ranges: Low range: 3 to 150 mg/L, meets ISO 15705:2002; middle-low range: 15 to 300 mg/L, meets ISO 15705:2002; middle range: 20 to 1500 mg/L, meets ISO 15705:2002; high range: 200 to 15000 mg/L
  • Applications: Raw water treatment, waste water treatment
  • Includes: Photometer, COD reactor RD 125, COD tube tests (0 to 150 mg/L and 0 to 1500 mg/L), tube stand, instruction manual, warranty information
  • CE Approved

Recently Viewed